AJ-2008B LCD Touch Screen Dialysis Machine
AJ-2008B
LCD Touch Screen Dialysis Machine
- Description
Description
Introduction
Size & Weight Size: 380mm×400mm×1380mm (L*W*H)
Area: 500*520 mm
Weight: 88KG
Power Supply AC220V, 50Hz / 60Hz, 10A
Input power: 1500W
Back-up battery: 30 minutes (optional)
Water input pressure: 0.15 MPa ~ 0.6 MPa, 21.75 P.S.I. ~ 87 P.S.I.
Water input temperature: 10℃ ~ 30℃
Working environment: temperature 10ºC ~ 30ºC at relative humidity of no more than 70%.
Dialysate
Dialysate temperature: preset range 34.0℃ ~ 39.0℃
Dialysate flux: 300~800 ml/min
Dialysate concentration: 12.1 mS/cm ~ 16.0 mS/cm, ±0.1 mS/cm
Dialysate mixing ratio:can set variety ratio.
UF rate Flow range: 0 ml/h ~ 4000 ml/h
Resolution ratio: 1ml
Precision: ±30 ml/h
Extracorporeal Part
Venous pressure: -180 mmHg ~ +600 mmHg, ±10 mmHg
Arterial pressure : -380 mmHg ~ +400 mmHg, ±10 mmHg
TMP pressure : -180 mmHg ~ +600 mmHg, ±20 mmHg
Blood pump flow range: 20 ml/min ~ 400 ml/min(diameter:Ф6 mm)
Spare pump flow range: 30 ml/min ~ 600 ml/min(diameter:Ф8 mm)
Resolution ratio: 1 ml
Precision: error range ±10ml or 10% of reading
Heparin Pump
Syringe size: 20, 30, 50 ml
Flow range: 0 ml/h ~ 10 ml/h
Resolution ratio: 0.1ml
Precision: ±5%
Sanitize
1.Hot decalcification :
Time: about 20 minutes
Temperature: 30~60℃, 500ml/min.
2.Chemical disinfection :
Time: about 45 minutes
Temperature: 30~40℃, 500ml/min.
- Heat disinfection :
Time: about 60 minutes
Temperature: >85℃, 300ml/min.
Storage Environment Storage temperature should be between 5℃ ~ 40℃, at relative humidity of no more than 80%.
Monitoring System
Dialysate temperature: preset range 34.0℃ ~ 39.0℃, ±0.5℃
Blood leak detection: Photochromic
Alarm when erythrocyte specific volume is 0.32±0.02 or blood leak volume is equal or more than 1ml per liter of dialysate
Bubble detection: ultrasonic
Alarm when a single air bubble volume is more than 200µl at 200ml/min blood flow
Conductivity: acoustic-optic, ±0.5%
Language
English and French etc.