Anesthesia Machine with Ventilator AJ-2105
Anesthesia Machine with Ventilator AJ-2105
(2 Vaporizers, 3 Gas)
Category: Operating Room Equipments
- Description
Description
Features
- The Ventilator adopts microcomputer control with high-definition 4” TFT color LCD display, integrating the functions of respiratory control and displaying the monitoring parameters. It is versatile in function and excellent in performance.
- The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose two kinds of gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.
- High-definition five-tube flowmeter (oxygen, nitrous oxide, and air), designed with stoppers for oxygen and nitrous oxide to control the output oxygen concentration.
- Reliable pneumatic transmission system.
- Economic and proper configuration of technologies, optional as required by the Customer.
- Modularized design and upgrading. Upgrading is available as required by the Customer.
TECHNICAL PARAMETERS
Part of Anesthesia Machine
- Main body High-strength engineering plastic rack, light,
- beautiful and corrosion resistant
- Scope of application adult, child
- Gas source O2: 0.27~0.55MPa N2O: 0.27~0.55MPa
- Flow meter O2 : 0.1~1.0L/min 1.1~10L/min
- N2O: 0.1~1.0L/min 1.1~10L/min
- Air : 0.1~10L/min
- O2, N2O linkage and N2O Stopper
- When using nitrous oxide, oxygen concentration>25%;
- When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
- Flow rate of rapid oxygen supply 25~75L/min
- Low oxygen pressure alarm There will be sound alarm when the oxygen pressure < 0.2MPa
- Vaporizer It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane,Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.
- Respiratory circuit working mode: all-close, semi-close, semi-open APL: 0.5~7 kPa
- Respiratory Bellows bellows for adults, tidal volume range:0~1500 ml bellows for children, tidal volume range:0~300 ml
- anyone can be chosen for application as
- required by the customer
Part of Ventilator
- Display mode High-definition 8.4” TFT color LCD screen display
- Ventilation mode IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH, IMV, VCV, PCV
- Ventilation parameters Tidal volume 20~1500 ml
- Rate 1~99 bpm
- SIMV rate 2~20 bpm
- I:E 4:1~1:8
- Inspiratory trigger pressure -1.0~2.0 kPa
- PEEP 0~2.0 kPa
- Pressure range 1.0~6.0 kPa
- SIGH 1.5 times the inspiratory time
- (70-150 adjustable)
- Inspiratory platform 0~50%
- Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure – time waveform, flow rate – time waveform,PEEP,inspiratory trigger pressure, Inspiratory platform, autonomous inspiratory rate, lung compliance, airway resistance, flow-volume loop, pressure-volume loop
- Monitoring of oxygen concentration 21%~100%
- Security Alarm System
- Oxygen Concentration alarm upper limit setting range 21%~100%
- low limit setting range 10%~80%
- Airway pressure alarm upper limit setting range 0.3~6.0 kPa
- low limit setting range 0.2~5.0 kPa
- Per-minute ventilation volume alarm upper limit setting range 3.0~30L/min
- low limit setting range 1.0~10L/min
- Sustained high-pressure alarm it will give alarm when stress have consistently
- been higher than 2.5 kPa
- Suffocation alarm it will give sound and light alarm if there
- is no tidal volume input for 15 seconds
- Power alarm
- Power AC 220V±10% 50Hz±1Hz
(UPS, Storage battery)
- (Mechanical Arms)
- (End-expiratory CO2 monitoring)
- Items within brackets are optional.