Our Products - Automatic Glycated Hemoglobin HbA1c Analyzer AC6601

Automatic Glycated Hemoglobin HbA1c Analyzer AC6601

Automatic Glycated Hemoglobin HbA1c Analyzer AC6601

Description

Basic Parameters
Testing Method Chromatography/ion exchange liquid chromatography
Testing Item Glycated hemoglobin HbA1c
Testing Scope 3% ~ 18%
Testing Parameters Precision (CV) ≤3%, Accuracy (V) ≤1.50%
Testing Time  4 minutes (including reporting)
Sample type Venous blood (EDTA anticoagulation),Peripheral finger blood
Sample volume 5μl whole blood for each sample
Functional Parameters
Photometer 415nm LED Integral flow colorimeter
Sampling mode Full-auto 25-position sample turntable (20 sample positions, 2 quality control positions, 1 emergency position, 1 cleaning position, 1 zero position)
Testing mode Auto test for batch, single test for emergency
Calibration Optional selection of 2 points or 4 points calibration, optional selection of manual or automation, Equipped with 1set of weight high and low level calibrators
Thermostatic control Column and reagent thermostatic control: 25 ± 0.5℃
Reagent matching One piece of column and reagent consumable/300 pieces of testing samples
Reagent allocation Each standard package includes A, B, C, D eluent, calibrator, hemolytic agent, pump tubing
Output Parameters
Display 320×240 LCD Graphic Display, displaying real-time testing curve
Printer 58mm Thermal printer, printing testing curve and report
Report Output IFCC concentration value, NGSP area percentage, ADAG average glucose
Data storage 1000 test report (including testing curve)
Communication interface RS232 communication interface, connecting to HIS/LIS system
Working Parameters
Power Supply AC220V/110V±22V  50/60Hz   215VA
Size 408mm×390mm×230mm
Weight  9Kg
Working Environment Temperature 12℃-30℃,Relative humidity ≤85%